Varex GS 3070 tube for Philips Tomoscan AV P3 systems
GS-3070 Rotating Anode X-Ray Tube
Scanner Types - Tomoscan SR5000
Housing Type - B-240H
OEM - Philips
The GS-3070 is a 5.5” (140 mm) 150 kV, 2.5 MJ (3.5 MHU) maximum anode heat content, rotating anode insert. This insert is specifically designed for Hitachi CT Scanners. The insert features a 7° tungstenrhenium facing on molybdenum with a graphite backed target and is available with the following nominal focal spot:
0.7 x 0.8
1.2 x 1.4
IEC 60336
Loading Factor for slit focal:
Small - 120 kV, 100 mA
Large - 120 kV, 200 mA
Maximum Anode Cooling Rate:
8,750 W (12,250 HU/sec)
Maximum continuous anode heat
dissipation:
3,400 W (4,760 HU/sec)
Nominal Anode Input Power:
Small - 23 kW IEC 60613
Large - 42 kW IEC 60613
Reference Axis:
Perpendicular to port face.
This insert is intended for use in Varex Imaging B-240H housing.